Measurement System Analysis (Gage R&R) Calculator

A gauge can pass on study variation and fail on tolerance. Both criteria matter.

Measurement Variation Standard deviations
units

One operator, one specimen, repeated readings

units

Spread between operators on the same specimen

Reference Scales Part and specification
units

True spread of the specimens in the study

units

Upper limit minus lower limit

Gauge R&R (% of Study Variation)

— %

Under 10% acceptable, 10–30% conditional, over 30% unacceptable (AIAG MSA)

Variance Breakdown

Combined Gauge R&R SD
— units
Total Study Variation SD
— units
Gauge R&R (% of Tolerance)
— %
Number of Distinct Categories (ndc)
—
Repeatability Share of Total
— %
Reproducibility Share of Total
— %

Variances add, standard deviations do not, so every combination here is a root-sum-square. Percent of study variation and percent of tolerance answer different questions and frequently disagree: the first depends on how representative the specimens were, so a study run on deliberately similar samples flatters the gauge and one run on extremes flatters it further. Percent of tolerance is the criterion that matters when the gauge is used to accept or reject against a specification, and it is independent of the specimens chosen. The ndc figure uses the AIAG 1.41 constant and should be at least 5; below 2 the gauge cannot distinguish parts at all and is only counting. None of this detects bias — a systematically wrong instrument can be perfectly repeatable and perfectly reproducible. Calibrate against a traceable standard as a separate exercise.

Measurement System Analysis (Gage R&R) Calculator — free, with the formula and a worked example, at Textile School.